HBM4EU ICI EQUAS report Chromium in serum round 2

  1. Home
  2. >
  3. Memphis Documents Posts
  4. >
  5. HBM4EU ICI EQUAS report Chromium in...
Download

136 Downloads

Last Updated: 17-10-2019 11:04

DescriptionPreviewVersions

Within the framework of the HBM4EU project, an Inter-Laboratory Comparison Investigation (ICI) was organized and conducted for the analysis of chromium in serum.
The study was performed from March 2019 to April 2019.
In total, 46 laboratories were invited for this second round, of which 22 laboratories from 17 countries registered.
The participation in this ICI was satisfactory; 21 out of 22 laboratories (95%) submitted their results.
In March 2019, two different test samples consisting of 3 mL serum spiked with chromium at two different concentrations (Crlow, Crhigh), one of each concentration, were prepared, and one sample of each concentration was sent to the participating laboratories for analysis.
Homogeneity and stability assessment of the control materials confirmed that the materials were adequately homogeneous and stable.
Laboratory results were rated using z-scores in accordance with ISO 13528 and ISO 17043. The standard deviation for proficiency assessment (also called target standard deviation) was set to FFP = 25%, as described in 5.3.
The evaluation of Crlow showed that 95% of the results were satisfactory, none were questionable and 5% were unsatisfactory. The evaluation of Crhigh showed that 100% of the results were satisfactory, none were questionable and none were unsatisfactory.