HBM4EU ICI EQUAS report Cadmium in urine round 2

  1. Home
  2. >
  3. Memphis Documents Posts
  4. >
  5. HBM4EU ICI EQUAS report Cadmium in...
Download

218 Downloads

Last Updated: 17-10-2019 11:01

DescriptionPreviewVersions

Within the framework of the HBM4EU project, an External Quality Assurance Scheme (EQUAS) was organized and conducted for the analysis of cadmium in urine.
The study was performed from October 2018 to January 2019.
The HBM4EU QAU had selected six expert laboratories for cadmium in urine. Four expert laboratories were from Europe (HBM4EU consortium) and three of them also participated as candidates. Two expert laboratories were from outside Europe (Japan, USA).
In total, 58 laboratories were invited for this second round, of which 39 laboratories from 22 countries registered.
The participation in this EQUAS was successful; 37 out of 39 laboratories (95%) submitted their results.
In November 2018, six different test samples consisting of 5 mL urine spiked with cadmium at two different concentrations (Cdlow, Cdhigh), three of each concentration, were prepared and sent to the participating expert and candidate laboratories for analysis.
Homogeneity and stability assessment of the control materials confirmed that the materials were adequately homogeneous and stable.
The expert assigned values were calculated by averaging the values obtained by the expert labs.
The proficiency of the laboratories was assessed through Z-scores calculated using the mean concentration as established by expert laboratories as assigned value, and a fixed fit-for-purpose relative target standard deviation (FFP-RSDR) of 25%.
The evaluation of Cdlow showed that 70% of the results were satisfactory, 3% were questionable and 27% were unsatisfactory. The evaluation of Cdhigh showed that 92% of the results were satisfactory, none were questionable and 8% were unsatisfactory.